Test-retest reliability of the KUDUwave™ for Standard and Extended High Frequencies. (Research report)

The test-retest reliability of all frequencies are within the required +/-10dB range. None of the test-retest results differed with more than 2.6dB on average. The test-retest reliability of all standard frequencies and of all extended high frequencies grouped together are 0.7dB and 1.4dB respectively. The test-retest reliability of the KUDUwave™at all frequencies are well within the acceptable range for accurate audiology testing.

Report generated by eMoyo with a large sample of randomised validation tests. Excellent data for validation.